Barium, Serum/Plasma Test (0519SP)

Barium, Serum/Plasma Test (0519SP)

Analysis Code 0519SP 
Test Name Barium, Serum/Plasma 
Test Includes Barium 
Purpose Exposure Monitoring/Investigation; This test is New York State approved. 
Category Metal/Element 
Method(s) Inductively Coupled Plasma/Mass
Spectrometry(ICP/MS) 
Specimen Requirements 1 mL Serum or Plasma 
Transport Temperature Refrigerated 
Specimen Container Plastic container (Acid washed or Trace metal-free) 
Special Handling Serum: Collect sample in Royal Blue top tube (Trace metal-free; No additive)
Plasma: Collect sample in Royal Blue top tube (Trace metal-free; EDTA)
Collect sample in plastic tube.
Promptly centrifuge and separate Serum or Plasma into an acid washed plastic screw capped vial using approved guidelines. 
Light Protection Required Not Required 
Stability Room Temperature: 30 day(s)
Refrigerated: 30 day(s)
Frozen (-20 °C): 30 day(s) 
*Rejection Criteria Glass container. Polymer gel separation tube (SST or PST). 
Day(s) Test Set-up / TAT [ICP/MS] Monday Wednesday Friday / 2 days 
Suggested CPT Code 83018 
**Minimum Volume 0.4 mL 