Titanium, Serum/Plasma Test (4486SP)

Titanium, Serum/Plasma Test (4486SP)

Analysis Code 4486SP 
Test Name Titanium, Serum/Plasma 
Test Includes Titanium 
Purpose Exposure Monitoring/Investigation; This test is New York State approved.

Note: On 06/14/2013 a change in the method for measuring titanium in serum has resulted in a shift with the reported findings. To compare a previous finding to a current result, multiply the old value by 1.44 and then subtract 66 for an approximate corresponding new value. The reporting limit for the new method is 10 mcg/L; therefore, any measured value less than 10 mcg/L will be reported as 'None Detected'. 
Category Metal/Element 
Method(s) Inductively Coupled Plasma/Mass
Spectrometry(ICP/MS) 
Specimen Requirements 2 mL Serum or Plasma 
Transport Temperature Refrigerated 
Specimen Container Plastic container (Acid washed or Trace metal-free) 
Special Handling Serum: Collect sample in Royal Blue top tube (Trace metal-free; No additive)
Plasma: Collect sample in Royal Blue top tube (Trace metal-free; EDTA)
Promptly centrifuge and separate Serum or Plasma into an acid washed plastic screw capped vial using approved guidelines. 
Light Protection Required Not Required 
Stability Room Temperature: 30 day(s)
Refrigerated: 30 day(s)
Frozen (-20 °C): 30 day(s) 
*Rejection Criteria Polymer gel separation tube (SST or PST). 
Day(s) Test Set-up / TAT [ICP/MS] Wednesday / 1 day 
Suggested CPT Code 83018 
**Minimum Volume 0.6 mL 
Test Summary Sheet Generate  

You are viewing Titanium, Serum/Plasma Test (4486SP)
*Rejection criteria pertain to clinical specimen submissions only.
**Stated minimum volume allows for a single analysis. Repeat analysis will not be performed.